RamSpec-VUV-0.3-0.6

VUV Raman Spectrometer


RamSpec-VUV is the first commercial Raman spectrometer available for the VUV region of the spectrum and the first instrument to measure Raman scattering signal in the VUV range of 177-190 nm.

The system has been designed to allow interchangeability between various grating arrangements for optimum configuration. Its relatively small size and standard connections also makes it an obvious choice for almost any system.



Specifications

Layout

  • Focal length: First and second stages: 304 mm,  Third stage: 608 mm
  • f / No. 6
  • Grating size: First and second stages: 50 mm, Third stage: 90 mm
  • Motor Drives: Stepper controlled sine drives with LabView interface
  • Wavelength  range:   140 nm  to 550 nm with standard grating
  • Wavelength accuracy:  0.03 nm
  • Wavelength reproducibility: 0.02 nm 
  • Focal plane: 7mm h x 30 mm w
  • Standard grating:  3600 G/mm
  • Standard CCD: Back-Thinned 28 mm 1044 pixels cooled to -45 C
  • Software: Labview CCD hyper-spectral display and scan control
  • Standard Slits: adjustable 25 to 4000 micron width
  • Resolution: 0.02 nm
  • Stray light:  <10(-8)
  • Vacuum:  <10(-5)
  • Dimensions: 1.1 m x 0.25 m x 0.7 m


 
 

 

 