- Focal
length:
305 mm
- f
/
No.
5.4
- Wavelength
range: 115
nm to 600 nm
- Wavelength
accuracy: +/- 0.3 nm with 3600 G/mm gtg
- Wavelength
reproducibility: +/- 0.1 nm
- Focal
plane: 30 mm maximum width
- Standard
grating: 3600 G/mm
- Grating
size: 50 x 50 mm
- Standard
CCD: Back Thinned 28 mm 2048 pixels
- Standard
Interchangeable
Slits: 25 to 4000 micron width
- Resolution:
0.03 nm with 25 um slits
- Stray
light: < 2 x 10 (-4) in VUV
- Vacuum:
<10(-6)
- Dimensions:
Main body 40 cm long. 18 cm w
and
h
|
|
- CCD
Detector
10
nm
to
1,000
nm
response 14 micron w by 200
micron h pixels
- XUV
back
thinned
Si
diode
1
nm
to 1,000 nm response
- CsTe
Photomultiplier
115
to
300
nm
response
- Gold
photo
diode
10
20
to
110
nm response
- CsI
photo
diode
115
to
170
nm
response
- CsTe
photo
diode
115
to
170
nm
response
- Windowless
light
sources
30
to
1,000
nm
- CsI
Micro
channel
plate
FO
coupled
CCD
10 to 170 nm
- RF
discharge
lamp
light
sources
115
to
1,000 nm
- Fiber
optic
couplings
- Motorized
wavelength
drive
with
low
cost
PC
control
- Labview
based
spectral
control
and
acquisition
software
|