VM600CT VUV/UV/Vis Monochromator and Spectrometer


 




VM600CT (background) pictured as a component of RamSpec-VUV Raman Spectrometer

  • High throughput f/5.4
  • Low astigmatism allows spatial as well as spectral imaging
  • Spectral range from 115 nm to 600 nm with 3600 G/mm grating
  • Holographic grating with high VUV efficiency
  • Can be easily switched from a monochromator to CCD-based spectrometer
  • Readily purged with Nitrogen or Helium for easy VUV
  • Can be pumped through entrance slit 
  • Back-Thinned CCD camera options
  • Fixed or Adjustable slits
  • Stepper motor drive with PC-based controller and Labview-based SW
  • Labview drivers and code for integration into user SW

The Resonance Ltd vacuum monochromator and spectrometer packs great performance in a small package. The system can be coupled directly to a vacuum chamber through a 2.75" Conflat® style flange to observe processes in a vacuum system.  The standard VUV grating has high dispersion and exhibits high efficiency in the VUV and UV  regions.  Low cost interchangeable gratings can be used to extend the range in the visible and IR regions. 

The system can be operated with a gas purge or evacuated allowing spectra to be obtained in the VUV  as well as regions in the IR where atmospheric absorption is high.

Specifications

 

Accessories

  • Focal length: 608 mm
  • f / No. 5.4
  • Wavelength  range:   115 nm  to 600 nm
  • Wavelength accuracy:  +/- 0.05 nm with 3600 G/mm gtg
  • Wavelength reproducibility:  +/- 0.03 nm 
  • Focal plane: 30 mm maximum width
  • Standard grating:  3600 G/mm
  • Grating size:  80 x 90 mm
  • Standard CCD: Back Thinned 28 mm 2048 pixels
  • Standard Interchangeable Slits: 25 to 4000 micron width
  • Resolution: 0.015 nm with 25 µm slits
  • Stray light: < 2 x 10 (-4) in VUV
  • Vacuum:  <10(-6)
  • Dimensions: 0.7 m x 0.35 m x 0.35 m
  • Motorized wavelength drive with low cost PC control
  • Labview based spectral control and acquisition software

  • CCD Detector  10 nm to 1,000  nm response 14 micron w by 200 micron h pixels
  • XUV back thinned Si diode 1 nm to 1,000 nm  response
  • CsTe Photomultiplier 115 to 300 nm response
  • Gold photo diode 10 20 to 110 nm response
  • CsI  photo diode  115 to 170 nm response
  • CsTe photo diode  115 to 170 nm response
  • Windowless light sources 30 to 1,000 nm
  • CsI Micro channel plate FO coupled CCD 10 to 170 nm
  • RF discharge lamp light sources 115 to 1,000 nm
  • Fiber optic couplings 

 
 

 

 

Copyright Resonance Ltd., October 2009
143 Ferndale Drive North, Barrie, ON L4N 9V9      Tel: 705-733-3633      Fax: 705-733-1388       Email: res@resonance.on.ca